Please use this identifier to cite or link to this item: https://idr.nitk.ac.in/jspui/handle/123456789/8171
Title: Handheld electrochemical workstation for serum albumin measurement
Authors: Hebbar, S.
Kumar, V.
Bhat, M.S.
Bhat, N.
Issue Date: 2016
Citation: 2016 IEEE International Conference on Distributed Computing, VLSI, Electrical Circuits and Robotics, DISCOVER 2016 - Proceedings, 2016, Vol., , pp.158-162
Abstract: This paper presents a novel handheld electrochemical workstation for serum albumin measurement. The system consists of a multi-path potentiostat module which performs electrochemical measurements on disposable test strip. The strip provides a port for applying blood sample. The test strip consists of 3 electrochemical cells for redundancy and parallel testing. The 3 sets of 3 electrodes (Working, Reference and Counter electrodes) are screen printed on a Polyethylene Terephthalate (PET) substrate. The system provides the unique capability of performing Cyclic Voltammetry and Chrono Amperometry measurements in three parallel paths. The system is very generic and flexible, with user defined inputs for voltage, sweep rate and time through 5 inch capacitive touch screen. The experimental results can be stored on micro SD card and the data is accessible with USB or Bluetooth interface. This paper reports an extensive characterization of system through several tests conducted on standard redox solution. The system is then validated for human serum albumin measurement, using clinical samples. This is the first ever point of care handheld diagnostic device in the world for human serum albumin measurement. � 2016 IEEE.
URI: http://idr.nitk.ac.in/jspui/handle/123456789/8171
Appears in Collections:2. Conference Papers

Files in This Item:
There are no files associated with this item.


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.