Automated low-temperature resistivity measurement setup: Design and fabrication

No Thumbnail Available

Date

2021

Journal Title

Journal ISSN

Volume Title

Publisher

Elsevier Ltd

Abstract

In this paper we present, the design and working of the automated low-temperature four-probe DC electrical resistivity measurement setup developed in-house. This setup measures the resistance/resistivity in the temperature range 300 K to 80 K. The overall arrangement comprises a low-temperature four-probe sample holder, Lakeshore 335 Temperature controller, Keithley Delta mode 6221 AC/DC source, 2182A Nano-voltmeter, and a vacuum system. For the measurement of sample temperature Pt-100 sensor is employed. The IEEE-488.2 card and GPIB cables link the mentioned instruments to the computer. The licenced Lab-VIEW 2018 software is employed to build the control programme. The established setup can measure the resistance of both bulk and thin-film samples, making it versatile. A test measurement on well-characterized ceramic Pr<inf>0.67</inf>Sr<inf>0.33</inf>MnO<inf>3</inf>perovskite oxide synthesized using a solid-state route has been carried out in the newly installed setup. The data acquired from the installed setup and beforehand resistance measurement carried out in-house low-temperature resistance measurement setup at UGC-DAE-Consortium for scientific research, Indore conceded a precision for both temperature and resistance. © 2021 Elsevier Ltd. All rights reserved.

Description

Keywords

Four-probe technique, LabVIEW, Low-temperature resistivity setup, Perovskite manganites

Citation

Materials Today: Proceedings, 2021, Vol.47, , p. 1670-1675

Endorsement

Review

Supplemented By

Referenced By